Scanning tunneling microscopy (STM) imaging requires a specially designed STM cartridge, which plugs into the Dimension Icon scanner and has a small tube for inserting an STM tip:
The holder contains a stainless steel tube with a 0.012" inner diameter mounted in ceramic at the center of the piezo tube. This design preserves the symmetry of the scan, although some may find it difficult to see the tip-sample interface. A circuit mounted on the probe holder contains the preamplifier for the tunneling current and provides connections to the tube electrodes.
The tip is tilted relative to the sample instead of pointing straight down. The tilt is designed to allow the user to see the tip in the optical view.
The STM cartridge is installed in the probe head in the same way as the AFM cartridge, with the tip pointing down. The instructions below explain how to insert STM tips into the STM cartridge.
To store an STM cartridge with a tip loaded, place the cartridge in a container with a lid with the sharp tip pointing up. Be sure to close the lid of the container.
To remove a tip from an STM cartridge, grip either end of the tip with needle-nose pliers, and pull the wire out of the hole.
www.bruker.com | Bruker Corporation |
www.brukerafmprobes.com | 112 Robin Hill Rd. |
nanoscaleworld.bruker-axs.com/nanoscaleworld/ | Santa Barbara, CA 93117 |
Customer Support: (800) 873-9750 | |
Copyright 2010, 2011. All Rights Reserved. |